EasyManuals Logo

ADVANTEST R3132 Series User Manual

ADVANTEST R3132 Series
588 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #217 background imageLoading...
Page #217 background image
R3132 Series Spectrum Analyzer Operation Manual
3.3.18 POWER MEASURE Key (Power Measurement)
3-62
ABS: Displays an absolute value using the unit selected from
the Units menu under the LEVEL key.
Parameter Setup Displays the Prmtr Setup menu.
Default Sets the following settings to default values: the frequency span,
resolution bandwidth, video bandwidth, sweep time, detector,
trace mode, CS/BS Setup,Nyquist Filter Setup and Limit Line
ON/OFF. When Default is enabled, the defaults are automatically
set when the spectrum mask is measured.
Manual Disables the Default function. As a result, the above settings must
be set manually. The currently set values are used when the spec-
trum mask is measured.
Define Default The currently set values become the new defaults.
Spectrum Mask OFF Turns the spectrum mask function off.
Spurious Freq The spurious table information is displayed after the Spuri Freq
menu has been displayed. (The spurious measurement table is not
displayed when there is no table data.)
When making spurious measurements while in frequency mode,
pass/fail judgment is repeated once for each frequency band after
the setup conditions have been selected and a maximum of 10 sig-
nal peaks have been searched for each frequency band. When this
function is used with Average Power measurement, pass/fail
judgments for average powers are performed.
Table No. 1/2/3 Sets Table 1,2 or 3.
1: Sets Table 1.
2: Sets Table 2.
3: Sets Table 3.
Edit Table Displays the Edit Table menu.
The editor used for the selected table number is displayed. You
can enter the start and stop frequencies, RBW, sweep time and
limit level of the frequency band which is used for pass/fail judg-
ment from this editor.
A maximum of 15 frequency bands can be set.
Table No. 1/2/3 Sets Table 1,2 or 3.
1: Sets Table 1.
2: Sets Table 2.
3: Sets Table 3.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3132 Series and is the answer not in the manual?

ADVANTEST R3132 Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3132 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals