5.2.15 Second Harmonic Distortion
5-58
23. On the R3162/72/82, set controls as follows:
Center Frequency: 1.9 GHz
Frequency Span: 1 kHz
24. On the R3162/72/82, press as follows to set fixed marker mode to on.
MKR, 1/2_more, Fixed MKR ON/OFF(ON)
25. On the R3162/72/82, set controls as follows:
Center Frequency: 3.8 GHz
Reference Level: -40 dBm
26. On the R3162/72/82, press as follows to set average mode for 20 samples.
TRACE, 1/2_more, AVG A, 2, 0, Hz(ENTER)
27. On the R3162/72/82, after average has completed, press PK SRCH to capture
signal peak.
28. Record the level of the delta marker reading on the performance verification
record sheet.