R3132 Series Spectrum Analyzer Operation Manual
5.2.16 Frequency Response
5-59
5.2.16 Frequency Response
The output of the signal generator is fed through a power splitter to a power sensor, then to the spectrum
analyzer. The signal generator’s power level is adjusted at 30 MHz to place the displayed signal at the cen-
ter horizontal graticule line of the spectrum analyzer. The power meter is placed in relative mode. At each
new signal generator frequency and spectrum analyzer center frequency, the signal generator’s power level
is adjusted to place the signal at the center horizontal graticule line. The power meter displays the inverse
of the frequency response relative to the signal of CAL OUT.
The R3132 series is phase locked to the signal generator’s 10 MHz reference.
Specification:
Apply for R3132
Preamplifier OFF, Attenuator 10 dB
±0.5 dB Frequency Range 100 kHz to 3 GHz (20°C to 30°C)
±2.0 dB Frequency Range 9 kHz to 3 GHz
Preamplifier ON, Attenuator 10 dB
±1.0 dB Frequency Range 100 kHz to 2.7 GHz
±2.0 dB Frequency Range 9 kHz to 3 GHz
Apply for R3132N
Preamplifier OFF, Attenuator 10 dB
±0.5 dB Frequency Range 100 kHz to 2.2 GHz (20°C to 30°C)
±2.0 dB Frequency Range 9 kHz to 2.2 GHz
Preamplifier ON, Attenuator 10 dB
±1.0 dB Frequency Range 100 kHz to 2.2 GHz
±2.0 dB Frequency Range 9 kHz to 2.2 GHz
Apply for R3162
Preamplifier OFF, Attenuator 10 dB
±0.5 dB Frequency Range 100 kHz to 3 GHz (20°C to 30°C)
±2.0 dB Frequency Range 9 kHz to 3.3 GHz
±2.0 dB Frequency Range 3.2 GHz to 8 GHz
Preamplifier ON, Attenuator 10 dB
±1.0 dB Frequency Range 100 kHz to 2.7 GHz
±2.0 dB Frequency Range 9 kHz to 3.3 GHz