R3132 Series Spectrum Analyzer Operation Manual
6.3.1 External Mixer Performance Verification Test Procedures
6-38
41. On the R3172/R3182, set control as follow;
Center Frequency: 70.1 GHz
42. On the SG, set control as follow;
Output Frequency: 70.1 GHz
43. On the R3172/3182, press SINGLE key for a sweep.
44. On the R3172/R3182, press PKSRCH key to capture the signal peak.
45. Record the level of peak marker in the measured value column on the perfor-
mance verification test data sheets.
46. Calculate actual value using following equation.
Actual Value = Abs (Calibrated value) - Abs (Measured value)
47. Record the result in the actual value column on the performance verification test
data sheets.
48. Repeat steps 41 through 47 for the frequency setting both of R3172/R3182 center
49. frequency and SG output frequency up-to 80 GHz step by 100 MHz.