EasyManua.ls Logo

Aeonmed Aeon8800A - Page 171

Aeonmed Aeon8800A
183 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
162
Warning: Do not user T-adapter with infants, as this adds 7ml dead space to the patient
ciruit.
Warning: Do not use the ISA gas analyzer with metered-dose inhalers or nebulized
medications as this may clog the bacteria filter.
Warning: Since a successful zeroing requires the presence of ambient Air(21% O
2
and 0%
CO
2
), ensure that the ISA is placed in a well ventilated place. Avoid breathing
near the ISA sidestream gas analyzer before or during the zeroing procedure.
Warning: Never sterilize or immerse the ISA sidestream gas analyzer in liquid.
Warning: The ISA sidestream gas analyzer is intended only as an adjunct in patient
assessment. It must be used in conjunction with other assessments of
clinical signs and symptoms.
Warning: Measurements can be affected by mobile and portable RF communications
equipment. Make sure that the ISA sidestream gas analyzer is used in the
electromagnetic environment specified in this manual.
Warning: Replace the sampling line if the sampling line input connector starts flashing
red, or the medical backboard device displays a “Check sampling line”
message.
Warning: No modification of this equipment is allowed without authorization of the
manufacturer. If this equipment is modified, appropriate inspection and testing
must be conducted to ensure continued safe operation.
Warning: The ISA sidestream gas analyzer is not designed are not designed for MRI
environments.
Warning: During MRI scanning, ISA must be placed outside the MRI suite.
Warning: Use of high frequency electrosurgical equipment in the vicinity of the
ISA/medical backboard device may produce interference and cause incorrect
measurements.
Warning: Do not apply negative pressure to remove condensed water from the
Nomoline Family sampling line.
Warning: Too strong positive or negative pressure in the patient circuit might affect the
sample flow.
Warning: Strong scavenging suction pressure might affect the sample flow.
Warning: Exhaust gases should be returned to the patient circuit or to a scavenging
system.
Warning: Due to the risk of patient cross-infection, always use a bacteria filter on the
exhaust port side if sampled gas is intended to be re-breathed.

Table of Contents