9.Technical data
163
Warning: Do not place the ISA gas analyzer in any position that might cause it to fall on
the patient.
Warning: Do not re-use disposable single-patient use Nomoline Family sampling line
due to the risk of cross contamination.
Warning: Do not sterilize or immerse Nomoline Family sampling line in liquid.
Warning: Do not operate the ISA sidestream gas analyzer if the enclosure is damaged.
Warning: Do not use the Nomoline Airway Adapter Set Infant with adult/pediatric
patients.
Caution: The ISA analyzers should be securely mounted in order to avoid the risk of
damage to the ISA.
Caution: Do not operate the ISA sidestream gas analyzer outside the specified
operating environment.
(US Only) Caution: Federal law restricts this device to sale by or on the order of a physician. For
professional use. See instructions for use full prescribing information, including indications
contraindications, warnings, precautions and adverse events.
Table 9-27
Model
ISA CO
2
/ AX+
Interface type
RS232, 9600bps
Accessories
Sampling tube, CO
2
, disposable.
Working principle of
gas module
The gas module is a sidestream gas monitoring module, connecting
to patient to monitor gas concentration of breath by electrical machine
inhaling trace gas.
The gas module can be used for adults, children and infants.
The gas module uses infrared technology, which determines gas
concentrations by measuring 10 different wavelengths according to
different absorption of gases to different ray.