2 m±0.1m and 2 m±0.1m versions
NOTE 1: After condensation, the unit shall be stored for more than 24 hours in an
environment with relative moisture content below 95%RH (non-condensing).
Table 9-29
ISA CO
2
: CO
2
ISA AX+: CO
2
, N
2
O, primary and secondary agents(HAL, ENF, ISO, SEV,
DES)
Automatic agent
identification
ISA AX+: Primary and secondary agent.
2-9 channel NDIR type gas analyzer measuring at 4~10μm.
When the detected value is offset;
When alarm occurs because “gas module monitor concentration
exceeding the limits”;
After replacing gas circuit adapter of gas monitor module.
ISA CO
2
: <10s(concentrations reported and full accuracy)
ISA AX+: <20s(concentrations reported, automatic agent identification
enabled and full accuracy)
Rise time2 at 50
ml/min sample flow
ISA CO
2
CO
2
≤200ms
ISA OR+/AX+
CO
2
≤300ms
N
2
O, O
2
, ENF, ISO, SEV, DES≤400ms
HAL≤500ms
Agent identification
time
Total system response
time
ISACO
2
< 3 seconds
ISA OR+/AX+CO
2
< 4 seconds
(with 2m Nomoline Airway Adapter Set sampling line)