166
The following accuracy specifications are valid for dry, single gases at 22±5 and 1013±40 hPa.
Table 9-31
Accuracy specifications - standard conditions
Gas
Range
Accuracy
CO
2
015%
1525%
± (0.2vol%+2% of reading)
Unspecified
N
2
O
0100%
± (2vol%+2% of reading)
HAL,ISO,ENF
08 %
825%
± (0.15vol%+5% of reading)
Unspecified
SEV
010%
1025%
± (0.15vol%+5% of reading)
Unspecified
DES
022%
2225%
± (0.15vol%+5% of reading)
Unspecified
Table 9-32
Accuracy specifications – all conditions
Gas
Accuracy
CO
2
± (0.3vol%+4% of reading)
N
2
O
± (2vol%+5% of reading)
Agents
1)
± (0.2vol%+10% of reading)
NOTE 1: The accuracy specification is not valid if more than two agents are present in
the gas mixture. If more than two agents are present, an alarm will be set.