1. If the E-flowmeter is working, the
message “Close the flowmeter” appears
on the user interface.
1) Select “Continue” to perform
Automatic Test again.
2) Select “Skip” to proceed to Manual
Test.
2. If the bellows is not full, the message
“Push “O
2
+” flush to completely fill the
Bellows” appears on the user interface.
1) Select “Continue” to perform
Manual Test again.
2) Select “Skip” to proceed to Manual
Test.
1. If the safety valve is valid and the
leakage is less than 100ml/min, the
automatic test is passed, and then the
system proceeds to the Manual Test.