62
2. If the leakage is in the range of 100
400ml/min, the interface shows as the
picture on the right.
1) Select “Retry” to perform the
Automatic Test again.
2) Select “Skip” to proceed to the
Manual Test, the automatic
ventilation is disabled.
3. If the leakage is more than 400ml/min,
the interface shows as the picture on the
right.
1) Select “Retry” to perform the
Automatic Test again.
2) Select “Skip” to proceed to the
Manual Test, the automatic
ventilation is disabled.
4. Automatic Test failed
If the safety valve in the gas module is
disabled the automatic test will fail, and
the interface as the picture on the right
appears.
Select “Continue” to perform Manual Test,
the automatic ventilation is disabled.