EasyManua.ls Logo

Agilent Technologies 1290 Infinity II 2D-LC

Agilent Technologies 1290 Infinity II 2D-LC
395 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4Installation
Hardware Installation
2D-LC User Guide 84
Figure 40 Setup G. Single Heart-Cutting Configuration as Single Sample Loop Setup
G
Number of
Capillary
# Connection ID x L [mm] P/N Description
1 1 Bypass capillary (OUT) 0.12 x 105 5500-1238 Capillary, ST 0.12x105 SL/SL
2 1 Waste line self-cut x 0.7 0890-1713 Tubing-flexible 0.8/1.61mm PTFE
WT (delivered with UV detector)
3 1 From pressure release kit; from
1
D column,
1
D detector
0.17 x 105
0.12 x 500
5500-1240
5500-1157
Capillary ST 0.17x105 SL/SL
Capillary ST 0.12x500 SL/S
4 Sample Loop (IN) 5004-0036 180 µL Loop 2D-LC as an example
5 Sample Loop (OUT) 5004-0036 180 µL Loop 2D-LC as an example
61
To
2
D column
0.12 x 400 5500-1251 Capillary ST 0.12x400 SL/SL
71
From
2
D pump
0.17 x 280 5067-4608 Capillary ST 0.17x280 SX/S
8 Bypass capillary (IN) 0.12 x 105 5500-1238 Capillary, ST 0.12x105 SL/SL
For all other capillaries / connections, see Figure 31 on page 73, Figure 32 on page 74, and Figure 33 on page 75.
NOTE
If the dual-loop setup has been selected in the software configuration (see “Configure the 2D-LC Cluster” on page 96),
install mirror-inverted, the sample loop at port 1 and 8 and the bypass capillary at position 4 and 5.

Table of Contents

Related product manuals