Maintenance and Troubleshooting 11
Essential Measurement Practices
1
Reagents and Reagent Mixes
Reagents and Reagent Mixes
Handle and store all reagents according to the instructions given in the
specific Kit Guide.
Keep all reagents and reagent mixes (for example, the gel-dye mix)
refrigerated at 4°C when not in use for more than 1 hour. Reagents left at
room temperature may decompose, leading to poor measurement results.
Allow all reagents and samples to equilibrate to room temperature for 30
minutes. Mix and spin down prior to use.
Gel and Gel-Dye Mix
Use gel-dye mix within the specified time frame stated in the instructions
from the kit guide. Otherwise, it may decompose and lead to poor
measurement results.
Protect dye and gel-dye mixes from light. Dye decomposes when exposed to
light.