Maintenance and Troubleshooting 133
Maintenance of the Chip Priming Station
9
Replacing the Syringe
Replacing the Syringe
1 Unscrew the old syringe from the top of the chip priming station.
2 Remove clip from the old syringe. Dispose syringe according to good
laboratory practices.
3 Slide new syringe into the clip. Ensure syringe and clip are flushed
together.
4 Screw the syringe tight into the luer lock adapter.
5 Check the priming station as described in “Checking the Chip Priming
Station for Proper Performance - Seal Test” on page 139.
When Quarterly or whenever it is clogged.
Parts required # p/n Description
1 NA Syringe kit that comes with each DNA, RNA and Protein kit
1 NA Deionized water