Maintenance and Troubleshooting 13
Essential Measurement Practices
1
Chips
Chips
Prepared chips must be used within 5 minutes. Reagents may evaporate,
leading to poor results.
For DNA and RNA assays, vortex chips for 1 minute. Inappropriate and
insufficient vortexing will lead to poor results. Use only the IKA vortexer
for chip vortexing. Replace the chip adapter if it is worn out. For the MS-2
vortexers with 3 mounting screws, the replacement part number is
5065-9966. For MS-3 vortexers with 4 mounting screws, replacement
adapters may be purchased directly from IKA (www.ika.de) with part
number 3428300.
Do not touch the wells of the chip. The chip could get contaminated
resulting in poor measurement results.
Do not leave any wells of the chip empty. The assay will not run properly.
For DNA and RNA assays: Add 1 μL of sample buffer to each unused sample
well so the total liquid volume in each well is at least 6 μL.
For protein assays: pipette a sample or ladder replicate in any empty
sample well.
Do not touch the underside of the chip.

Table of Contents

Other manuals for Agilent Technologies 2100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 2100 and is the answer not in the manual?

Agilent Technologies 2100 Specifications

General IconGeneral
BrandAgilent Technologies
Model2100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals