136 Maintenance and Troubleshooting
9
Maintenance of the Chip Priming Station
Replacing the Syringe Adapter
Replacing the Syringe Adapter
1 Follow the steps described in “Cleaning the Syringe Adapter” on page 134
to remove the syringe adapter.
2 Dispose the old syringe adapter.
3 Insert the syringe adapter, see Figure 5 on page 135.
4 Follow the steps as described in “Cleaning the Syringe Adapter” on
page 134 to reassemble the priming station.
5 Check the priming station as described in “Checking the Chip Priming
Station for Proper Performance - Seal Test” on page 139.
When If significantly clogged and unable to clean thoroughly.
Parts required # p/n Description
1 G2938-68716 Gasket kit