Maintenance and Troubleshooting 119
Maintenance of the Electrode Cartridge
8
RNA Nano Assay
RNA Nano Assay
Assay Before each run After each run Monthly or after liquid spill: Pin
set cleaning
RNA Nano Electrode cleaner:
RNase ZAP for 60 seconds.
RNase free H
2
O for
10 seconds.
Electrode cleaner:
RNase free H
2
O for
10 seconds.
RNase decontamination with
brush:
RNase ZAP.
RNase free H
2
O.