14 Maintenance and Troubleshooting
1
Essential Measurement Practices
Agilent 2100 Bioanalyzer System
Agilent 2100 Bioanalyzer System
Do not touch the 2100 Bioanalyzer instrument during a run and never place
it on a vibrating surface or near air-circulating instruments (for example,
temperature cyclers).
Do not force the chip to fit in the 2100 Bioanalyzer instrument. The
electrode cartridge may be damaged when the lid is closed.
Clean electrodes on a daily basis using the electrode cleaner. For more
details, see “Maintenance of the Electrode Cartridge” on page 115.
Thorougly clean electrodes on a monthly basis using a toothbrush and
distilled water. For more details, see “How to Clean the Pin Set of the
Electrode Cartridge” on page 126.
Clean the focusing lens once a month (or after any liquid spill) using
isopropanol. For more details, see “Cleaning the Lens” on page 143.