Maintenance and Troubleshooting 123
Maintenance of the Electrode Cartridge
8
RNA Pico or Small RNA Assay
RNA Pico or Small RNA Assay
Assay Before each run After each run Monthly or after liquid spill:
pin set cleaning
RNA Pico or
Small RNA
Electrode cleaner:
RNase free H
2
O for
5minutes.
Electrode cleaner:
RNase free H
2
O for
30 seconds.
RNase decontamination with
brush:
RNaseZAP
RNase free H
2
O