Maintenance and Troubleshooting 63
Troubleshooting the DNA Application
5
Symptoms (High Sensitivity DNA)
Artefact Peaks
Back to “Symptoms (High Sensitivity DNA)” on page 62
Most probable causes Solution
Samples are in water. Refer to the Kit Guide for sample buffer specifications. For optimal results,
samples should be dissolved in 10 mM Tris and 1 mM EDTA.
Probable causes Solution
Chip, gel-dye mix, or samples are
contaminated with particles.
Wear powder-free gloves only.
Do not touch the wells of the chip.
Clean the electrodes.
Load the chip immediately after taking it out of its sealed bag.