Maintenance and Troubleshooting 65
Troubleshooting the DNA Application
5
Symptoms (High Sensitivity DNA)
Baseline Negative Dips
Most probable causes Solution
Residual RNase ZAP and/or SDS
contamination on the electrode pins.
Only use RNaseZAP when decontaminating the
pins according to “How to Clean the Pin Set of
the Electrode Cartridge” on page 126.
Clean the pin set thoroughly with water after
running Protein or RNA assays. Change the
water in the electrode cleaner Chip and/or the
electrode cleaner Chip itself regularly.