150 Maintenance and Troubleshooting
11
Maintenance of the Vortexer
Changing the Adapter
Changing the Adapter
1 Release the 3 or 4 screws on top of the adapter.
2 Hold the base of the vortex mixer and pull up the head. Discard the old
head according to good laboratory practices.
3 Place the new head adapter on the vortex mixer.
4 Insert and fix the 3 or 4 screws with the screw driver.
When Whenever the vortex adapter is damaged.
Tools required p/n Description
NA Screw driver
Parts required # p/n Description
1 5065-9966 IKA vortex mixer adapter (for MS 2 vortexers)
1 IKA 3617036 IKA vortex mixer adapter (for MS 3 vortexers) NOTE: This part must be
purchased directly from IKA.