EasyManua.ls Logo

Agilent Technologies 2100 - Checking the Chip Priming Station for Proper Performance - Seal Test

Agilent Technologies 2100
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Troubleshooting 139
Maintenance of the Chip Priming Station
9
Checking the Chip Priming Station for Proper Performance - Seal Test
Checking the Chip Priming Station for Proper Performance - Seal
Test
1 Make sure the syringe is tightly connected to the chip priming station.
2 Pull the plunger of the syringe to the 1.0 mL position (plunger pulled back).
3 Place an unused chip in the chip priming station.
4 Close the chip priming station. The lock of the latch will audibly click when
it closes.
5 Press the plunger down until it is locked by the clip. This is shown in
Figure 8 on page 139.
Figure 8 Locking the plunger of the syringe with the clip
When Every month or whenever a component of the priming station (syringe, adapter or gasket) was
replaced.
Parts required Description
Unused DNA or RNA Chip
Plunger
Clip holding
plunger in place

Table of Contents

Other manuals for Agilent Technologies 2100

Related product manuals