140 Maintenance and Troubleshooting
9
Maintenance of the Chip Priming Station
Checking the Chip Priming Station for Proper Performance - Seal Test
6 Wait for 5 seconds and lower latch of the clip to release the plunger as
shown in Figure 9 on page 140.
Figure 9 Releasing the plunger from the syringe
7 To indicate an appropriate sealing, the plunger should move back up at
least to the 0.3 mL mark within less than 1 second.
If the plunger does not move up to the 0.3 mL mark within a second, the syringe-chip
connection is probably not tight enough. Retighten the syringe or replace the syringe
adapter, syringe or gasket to fix the problem.