EasyManua.ls Logo

Agilent Technologies 2100 - Page 140

Agilent Technologies 2100
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
140 Maintenance and Troubleshooting
9
Maintenance of the Chip Priming Station
Checking the Chip Priming Station for Proper Performance - Seal Test
6 Wait for 5 seconds and lower latch of the clip to release the plunger as
shown in Figure 9 on page 140.
Figure 9 Releasing the plunger from the syringe
7 To indicate an appropriate sealing, the plunger should move back up at
least to the 0.3 mL mark within less than 1 second.
Lower latch
Syringe clip
NOTE
If the plunger does not move up to the 0.3 mL mark within a second, the syringe-chip
connection is probably not tight enough. Retighten the syringe or replace the syringe
adapter, syringe or gasket to fix the problem.

Table of Contents

Other manuals for Agilent Technologies 2100

Related product manuals