56 Maintenance and Troubleshooting
5
Troubleshooting the DNA Application
Symptoms (DNA)
Peak Tailing
Back to “Symptoms (DNA)” on page 37
Most probable causes Solution
Sample salt concentration is too high. Refer to the maximum sample buffer salt limits as specified in the Kit Guide.
Dilute samples with deionized DNase free water, if necessary.