Maintenance and Troubleshooting 135
Maintenance of the Chip Priming Station
9
Cleaning the Syringe Adapter
3
Press the syringe adapter out of its mount as shown in
Figure 5 on page 135
.
Figure 5 Removing/replacing the syringe adapter
4 Remove dried gel at the opening of the adapter with a needle.
5 Screw on syringe and flush water through the adapter several times.
6 Flush syringe with isopropanol.
7 Allow adapter to dry fully.
8 Insert the syringe adapter, see Figure 5 on page 135.
9 Follow the steps as described in “Cleaning the Syringe Adapter” on
page 134 to reassemble the priming station.
10 Close the chip priming station.
11 Screw a dry syringe tight into the luer lock adapter.
12 Check the priming station as described in “Checking the Chip Priming
Station for Proper Performance - Seal Test” on page 139.