EasyManua.ls Logo

Agilent Technologies 2100 - Page 111

Agilent Technologies 2100
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Troubleshooting 111
Troubleshooting the Protein Application
7
Symptoms (Protein)
Error Message: No data received since 5 seconds
Most probable causes Solution
Disrupted communication between
instrument and computer.
Please refer to “Troubleshooting the Instrument Communication” on page 15 for
troubleshooting instrument communication issue.
Ensure the Agilent USB-Serial Adapter cable, black cable (5188-8031) for 2100
Expert Software version B.02.08 and greater is used to connect the 2100
Bioanalyzer instrument to the computer through a USB port. See “USB to Serial
Adapter” on page 21.

Table of Contents

Other manuals for Agilent Technologies 2100

Related product manuals