Save/Recall (Setups, Screens, Data) 18
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 279
2 Press the softkey in the second position and use the Entry knob to
navigate to the save location. See "To navigate storage locations" on
page 282.
3 Press the Settings softkey.
In the File Settings Menu, you have these softkeys and options:
Setup Info — when enabled, setup information (vertical, horizontal,
trigger, acquisition, math, and display settings) is also saved in a
separate file with a TXT extension.
Length — sets the number of data points that will be output to the
file. For more information, see "Length Control" on page 279.
Save Seg — when data is acquired to segmented memory, you can
specify whether the currently displayed segment is saved or all
acquired segments are saved. (See also "Saving Data from Segmented
Memory" on page 202.)
4 Finally, press the Press to Save softkey.
A message indicating whether the save was successful is displayed.
See Also "Binary Data (.bin) Format" on page 342
"CSV and ASCII XY files" on page 348
"Minimum and Maximum Values in CSV Files" on page 349
Length Control
The Length control is available when saving data to CSV, ASCII XY, or BIN
format files. It sets the number of data points that will be output to the
file. Only displayed data points are saved.
The maximum number of data points depends on these things:
Whether acquisitions are running. When stopped, data comes from the
raw acquisition record. When running, data comes from the smaller
measurement record.
Whether the oscilloscope was stopped using [Stop] or [Single]. Running
acquisitions split memory to provide fast waveform update rates. Single
acquisitions use full memory.
Whether only one channel of a pair is turned on. (Channels 1 and 2 are
one pair, channels 3 and 4 are the other.) Acquisition memory is
divided among the channels in a pair.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals