194 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
12 Acquisition Control
• Normal — at slower time/div settings, normal decimation occurs, and
there is no averaging. Use this mode for most waveforms. See
"Normal Acquisition Mode" on page 194.
• Peak Detect — at slower time/div settings, the maximum and minimum
samples in the effective sample period are stored. Use this mode for
displaying narrow pulses that occur infrequently. See "Peak Detect
Acquisition Mode" on page 194.
• Averaging — at all time/div settings, the specified number of triggers
are averaged together. Use this mode for reducing noise and
increasing resolution of periodic signals without bandwidth or rise
time degradation. See "Averaging Acquisition Mode" on page 197.
• High Resolution — at slower time/div settings, all samples in the
effective sample period are averaged and the average value is stored.
Use this mode for reducing random noise. See "High Resolution
Acquisition Mode" on page 199.
Normal Acquisition Mode
In Normal mode at slower time/div settings, extra samples are decimated
(in other words, some are thrown away). This mode yields the best display
for most waveforms.
Peak Detect Acquisition Mode
In Peak Detect mode at slower time/div settings, minimum and maximum
samples are kept in order to capture infrequent and narrow events (at the
expense of exaggerating any noise). This mode displays all pulses that are
at least as wide as the sample period.
For InfiniiVision 3000 X-Series oscilloscopes, which have a maximum
sample rate of 4 GSa/s, a sample is taken every 250 ps (sample period).
See Also • "Glitch or Narrow Pulse Capture" on page 195
• "Using Peak Detect Mode to Find a Glitch" on page 196