377
Agilent InfiniiVision 3000 X-Series Oscilloscopes
User's Guide
s1
25
I2C/SPI Triggering and Serial Decode
Setup for I2C Signals 377
I2C Triggering 378
I2C Serial Decode 382
Setup for SPI Signals 386
SPI Triggering 389
SPI Serial Decode 391
I2C/SPI triggering and serial decode requires Option EMBD or the
DSOX3EMBD upgrade.
Setup for I2C Signals
I
2
C (Inter-IC bus) signals setup consists of connecting the oscilloscope to
the serial data (SDA) line and the serial clock (SCL) line and then
specifying the input signal threshold voltage levels.
To set up the oscilloscope to capture I
2
C signals, use the Signals softkey
which appears in the Serial Decode Menu:
1 Press [Label] to turn on labels.
2 Press [Serial].
3 Press the Serial softkey, turn the Entry knob to select the desired slot
(Serial 1 or Serial 2), and press the softkey again to enable decode.
NOTE
Only one SPI serial bus can be decoded at a time.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals