298 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
20 Utility Settings
 
3 Press the LAN Settings softkey.
4 Press the Config softkey; then, turn the Entry knob to select Automatic, 
and press the softkey again to disable it.
If Automatic is not enabled, the oscilloscope's LAN configuration must 
be set up manually using the Addresses and Host name softkeys
5 Configure the oscilloscope's LAN interface:
a Press the Addresses softkey.
b Use the Modify softkey (and the other softkeys and the Entry knob) 
to enter the IP Address, Subnet Mask, Gateway IP, and DNS IP 
values. When you are done, go back up in the menu hierarchy.
c Press the Host name softkey. Use the softkeys and the Entry knob to 
enter the Host name. When you are done, go back up in the menu 
hierarchy.
d Press the Apply softkey.
6 Connect the oscilloscope to the local area network (LAN) by inserting 
the LAN cable into the "LAN" port on the rear panel of the 
oscilloscope.
Stand-alone (Point-to-Point) Connection to a PC
The following procedure describes how to establish a point-to- point (stand 
alone) connection to the oscilloscope. This is useful if you want to control 
the oscilloscope using a laptop computer or a stand- alone computer.
1 Press  [Utility] > I/O.
2 Press the LAN Settings softkey.
3 Press the Config softkey; then, turn the Entry knob to select Automatic, 
and press the softkey again to enable it.
If your network supports DHCP or AutoIP, enabling Automatic lets the 
oscilloscope use those services to get its LAN configuration settings
4 Connect your PC to the oscilloscope using a cross- over LAN cable such 
as Agilent part number 5061- 0701, available on the web at 
"www.parts.agilent.com".
5 Cycle power on the oscilloscope. Wait until the LAN connection is 
configured:
• Press  [Utility] > I/O and wait until the LAN status shows "configured".