150 BD FACSCalibur Instructions for Use
Preparing an Acquisition Document
For acquisition, Worklist Manager requires access to an application-specific
document with defined acquisition conditions. When you click Run Tests at the
Worklist view, Worklist Manager uses the defined document to launch the
acquisition software.
BD CellQuest Pro
BD Multiset BD HLA-B27
1 Define an Experiment
document with
appropriate Acquisition
& Storage conditions.
; Use Acquisition-to-
Analysis plots to
print data after each
tube is acquired
; For Collection
Criteria, use Event
Count or Time to
ensure you will not
run out of sample.
2 Define a panel.
3 Select the panel in the
Acquisition Browser.
4 Save the document in BD
Applications > Worklist
Manager > CellQuest
Experiments Folder.
1 Use the default Schedule
document or define a
Schedule document with
choices from Test Prefs.
NOTICE For absolute
counts, do not mix lots
of BD Trucount tubes in
the same worklist. Make
sure the BD Trucount
bead information in the
Schedule document is
correct before you run it
with Worklist Manager
software.
2 Customize Preferences.
3 Save the document
where you want.
1 Use the default Schedule
document or define a
Schedule document with
your preferred number
of events to acquire.
2 Customize Preferences.
3 Save the document
where you want.