184 BD FACSCalibur Instructions for Use
5 Install the new seal.
The Loader accessory kit includes 3 versions of the seal—thick, medium,
and thin. Install the next-thickest version, compared to the seal that is
currently installed. Make sure it is flush against the SIT.
For reordering information, see page 213.
6 Replace the outer droplet sleeve and tighten the SIP retainer.
7 Replace the tube guide retainer and the tube guide.
8 Turn on the instrument main power, and test the new seal by loading a tube
with the Lifter key on the Loader keypad.
Cleaning the Tube Lifter
The Loader tube lifter might need cleaning if a vertical timeout error is reported
by Loader Manager software. Use the Single Commands dialog to load a single
tube so you can access the tube lifter.
1 Place an empty test tube in position 1 on a Loader rack, and install the rack
on the Loader.
2 Push in the Loader drawer and replace the cover.
3 Confirm that all software applications are closed.
4 Launch Loader Manager, and click Maintenance.
5 Click Single Commands in the Maintenance and Diagnostics window, and
click Run.