182 BD FACSCalibur Instructions for Use
5 Verify Print Port (2) is selected; click Save.
If you continue to have connectivity problems after checking all cable
connections, contact your BD Biosciences service representative.
Replacing the Loader Seal
The Loader seal can become worn over time and need replacement. If you have
trouble with tube pressurization errors, first see page 208. If no other solutions
work, replace the seal.
1 Turn off the instrument main power.
2 Remove the tube guide retainer and attached tube guide.
Turn the tube guide and retainer counterclockwise, as shown in Figure 8-5.
Any instrument surface that comes in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when cleaning
the instrument or replacing parts. Wear suitable protective clothing,
eyewear, and gloves.