Chapter 2: BD FACStation Software 47
BD HLA-B27 Software
Use BD HLA-B27 software to classify a sample as BD HLA-B27 negative or
positive. This classification is based on the measured mean fluorescence intensity
(MFI) compared to a predetermined decision-marker setting encoded in the suffix
of the reagent antibody lot number.
Figure 2-5 BD HLA-B27 application window
The icon bar shows where you are in the run—the icon for the current view is
highlighted. Do the following at each view.
Sign in as the operator.
Define run-specific information.