Chapter 6: Running Samples with the Loader 163
Pausing Acquisition
NOTICE To interrupt the Loader during a BD Multiset assay, pause in the
BD Multiset Acquisition or Lab Report view. Do not use Worklist Manager to
stop a BD Multiset run; do not stop by removing the Loader cover.
NOTICE After an aborted run, the long clean is not performed automatically.
You will then have to run cleaning manually (page 166).
Do the following to stop the Loader during a BD CellQuest Pro assay.
1 Click Stop at the Worklist view.
2 When the countdown dialog appears, click Yes.
3 Choose from the following:
Rerun—reruns the current tube
Next Tube—skips to the next tube on the Worklist
Next Sample—advances the Loader to the first tube of the next sample
(if available), leaving the previous sample incomplete
Stop Run—advances to the Summary Report view