56 BD FACSCalibur Instructions for Use
Filling the Sheath Reservoir
Figure 3-1 Fluidics drawer
1 Slide out the fluidics drawer.
2 Push STNDBY and push the vent valve toward the arrow to depressurize
the system.
3 Disconnect the sheath tubing (white) and the air supply tubing (blue) from
the instrument.
4 Disconnect the sheath level sensor line.
5 Slide the metal bracket away from you, and lift up to remove it.
6 Remove the sheath reservoir.
7 Unscrew the cap assembly and remove it.
8 Fill the reservoir to 75% capacity.
See page 214 for recommended fluids
.
vent valve
level sensor
sheath reservoir
sheath tubing
air supply tubing
ball valve tab
metal bracket
line