210 BD FACSCalibur Instructions for Use
Worklist Manager Troubleshooting
Observation Possible Causes Recommended Solutions
BD CellQuest Pro
panel unavailable
Experiment document in
wrong folder
Place Experiment document(s)
in BD Applications > Worklist
Manager > CellQuest
Experiments Folder.
No panel selected Open document in
BD CellQuest Pro software and
select panel in the Acquisition
Browser.
BD Multiset assay
unavailable
PanelReagentFile alias
missing from ClinApps Info
Folder or corrupted
• Make alias of
PanelReagentFile and place
in BD Applications >
ClinApps Info Folder.
• Re-install BD Multiset
software.
Unlisted instrument
settings file
Instrument settings in
wrong folder
Move the file to BD Files >
Instrument Settings Files.