222 BD FACSCalibur Instructions for Use
Signal Processing
Loader Specifications
Workstation resolution 1,024 channels on all parameters
Data acquisition channels Logarithmic amplifiers for SSC, FL1, FL2, FL3, and
FL4 (with FL4 option), providing four-log decade
range
Fluorescence compensation Fluorescence spectral overlap can be compensated
between FL1 and FL2, FL2 and FL3, and FL3 and
FL4 (with FL4 option)
Pulse processing Width and area measurements for discriminating
doublets; available for all fluorescence parameters
Time Can be correlated to any parameter
Rack compatibility Loader carousels, numbers 1–16
Tube compatibility
Carousel Up to 40 uncapped 12 x 75-mm tubes
BD Falcon polystyrene test tubes
BD Trucount tubes
Thickness of accumulated labels 5 mils (127 μm)
Tube sample volume <1 mL