Maintenance
Scheduled Maintenance on page 170
Unscheduled Maintenance on page 175
Loader Maintenance on page 179
Cleaning Solution
For cleaning, use one of the following:
BD FACS Clean (For ordering information, see page 215.)
10% bleach (1:10 dilution of household bleach in DI water)
0.5% sodium hypochlorite
Any instrument surface that comes in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when cleaning
the instrument or replacing parts. Wear suitable protective clothing,
eyewear, and gloves.
Higher concentrations of sodium hypochlorite and other cleaning agents
can damage the instrument.