180 BD FACSCalibur Instructions for Use
Checking Tube Lifter Calibration
Tube lifter parameters are used to adjust the tube lifter to ensure proper sealing
with the SIP. You might be asked to locate these settings for BD Biosciences
personnel during troubleshooting.
1 Quit Worklist Manager, if running, and launch Loader Manager.
2 Click Maintenance.
3 In Maintenance and Diagnostics, select Check Lifter Parameters, and click
Run. Parameters show in a separate window:
4 (Optional) Click Print.
5 Click Done, Done, and close the Loader Status window.