Chapter 8: Maintenance 185
6 Select the Park lifter command in the left column, and click Send.
Ensure the tube number is 1 in the Enter the tube number to load field.
7 Once the lifter is raised, switch off the Loader power on the keypad.
8 Remove the Loader cover.
9 Hold the test tube with one hand while you pull out the Loader drawer
with the other hand.
10 Remove the Loader rack and test tube, and place the test tube in the rack.
11 Use a gauze pad and DI water to wipe the tube lifter shaft and base.
12 Push in the Loader drawer and replace the cover.
13 Switch on the Loader power.
The lifter will automatically lower.
14 Select the Reset controller command in the Single Commands dialog, and
click Send.
15 Install a rack with a tube in position 1 and check the tube lifting and sealing.
; Tip Place the cytometer in RUN.