BD FocalPoint™ GS Imaging System Instrument User’s Manual
202
The following table lists the Cautions and Warnings found in this manual.
Notice
Type
Located on Found on Notice
Warning Page 23 BD FocalPoint™
GS Review
Station
Make sure that the clip grips the slide edge securely. An unsecured
slide could fall into a slotted stage opening and be damaged.
Damaged or broken slides can cause injuries. Be careful not to allow
the clip to snap against the glass.
Caution Page 32 BD FocalPoint™
GS Review
Station
All FOVs must be reviewed before performing full slide review.
Caution Page 50 BD FocalPoint™
GS Review
Station
This feature provides the history of BD FocalPoint™ GS Review
Station slide information only, not the patient history.
Caution Page 60 BD FocalPoint™
GS Review
Station
When the server is configured for automatic data transfer, do not
perform a manual data transfer.
Caution Page 60 BD FocalPoint™
GS Review
Station
Your images are stored on the BD FocalPoint™ Slide Profiler for 10
days. Your copy service settings should be set to less than 10 day
intervals.
Warning Page 137 BD FocalPoint™
Slide Profiler
Shock and fire potential from spilled liquids. Do not place containers
with liquids on the device or the workstation cart. Do not spill liquids
on the system; fluid seepage into internal components creates a
potential shock and fire hazard. Shut down the device, disconnect
from the power source and wipe up all spills immediately. Do not
operate the system if internal components have been exposed to
fluid.
Warning Page 138 BD FocalPoint™
Slide Profiler
Be careful not to drop or break slides during slide preparation and
when loading and unloading slides into trays. Broken slides may
cause injuries.
Warning Page 138 BD FocalPoint™
Slide Profiler
Electromagnetic fields. This is a Class A product. In a domestic
environment, this product may cause radio interference with other
electronic devices, such as telephones and other medical equipment,
in which case the user may be required to take measures to reduce
such interference.
Warning Page 141 BD FocalPoint™
Slide Profiler
Remove all obstructive jewelry before loading or unloading trays.
After opening a hopper door, be sure all moving parts in the hopper
have stopped before inserting or removing a tray. If trays are inserted
before all moving parts have stopped, injury may result or the
instrument may malfunction.
Caution Page 141 BD FocalPoint™
Slide Profiler
If slides that do not match the operating mode of the instrument are
successfully processed on the instrument, the slides results reporting
for the print set will be invalid. All the slides in the print set must be
rerun on an instrument with the correct operating mode.
Caution Page 142 BD FocalPoint™
Slide Profiler
When adding trays to the input hopper during processing, be careful
not to touch the extended edge of a tray currently being processed.
Touching the tray could invalidate the positioning determinations
made for the cellular features on the slide in process.
Caution Page 153 BD FocalPoint™
Slide Profiler
Contact BD Technical Support before attempting to shut down the
instrument.
Follow the shutdown procedure exactly.