Warnings and Cautions
203
Table 22-2 – Cautions and Warnings
Caution Page 154 BD FocalPoint™
Slide Profiler
It is important to power down the system components exactly in the
order shown below:
Caution Page 158 BD FocalPoint™
Slide Profiler
The BD FocalPoint™ GS Imaging System contains no user-
replaceable or serviceable parts. When cleaning the calibration plate,
be careful not to disturb any internal components.
Warning Page 159 BD FocalPoint™
Slide Profiler
Shock potential when cleaning with power applied. Always turn off
the power switch and unplug the power cord before cleaning the
internal components of the device.
Warning Page 160 BD FocalPoint™
Slide Profiler
Shock and fire potential. Failure to unplug the monitor from the power
socket before performing this procedure could result in an electric
shock or fire.
Caution Page 161 BD FocalPoint™
Slide Profiler
Do not clean the instrument while it is processing slides.
Warning Page 161 BD FocalPoint™
Slide Profiler
Shock and fire potential when power applied improperly. The symbol
next to the power connector indicates potential shock hazard. Ensure
that the system is connected to a power receptacle that provides
voltage and current within the specified rating for the system. Use of
an incompatible power receptacle may produce electrical shock and
fire hazards.
Warning Page 161 BD FocalPoint™
Slide Profiler
Shock and fire potential when improperly grounded. Never use a
two-prong plug adapter to connect primary power to the system. Use
of a two-prong adapter disconnects the utility ground, creating a
potential shock hazard. Always connect the system power cord
directly to an appropriate receptacle with a functional ground.
Warning Page 137 BD FocalPoint™
Slide Profile
The BD FocalPoint™ Slide Profiler must be positioned in such a
manner that both, the power switch and the AC input are easily
accessible to the operator at all times.Unplug the supply cord from
the main circuit to disconnect the device.
Warning Page 137 BD FocalPoint™
Slide Profile
Protection provided by this equipment may be impaired if it is used in
a manner not consistent with the instructions in this manual.
Warning Page 137 BD FocalPoint™
Slide Profile
The FocalPoint™ Slide Profiler must be grounded at all times. To
ensure that the power cord and outlet strip have the required
electrical current ratings, only use the cord and power strip supplied
with the unit, or as provided by BD.
Warning Page 137 BD FocalPoint™
Slide Profiler
Shock and fire potential when improperly grounded. Never use a
two-prong plug adapter to connect primary power to the system. Use
of a two-prong adapter disconnects the utility ground, creating a
potential shock hazard. Always connect the system power cord
directly to an appropriate receptacle with a functional ground.
Caution Page 162 BD FocalPoint™
Slide Profiler
When performing this procedure, be careful not to touch or depress
any of the mouse buttons. Doing so could cause unexpected results,
as the mouse is “live” and the pointer is wandering around the
screen.
Caution Page 213 BD FocalPoint™
Slide Profiler
Only a BD-authorized representative should change the barcode
reader format. If the barcode reader format needs to be changed,
please contact your authorized service representative for
instructions.
Notice
Type
Located on Found on Notice

Table of Contents