98 MC6 User Manual – Appendix
THE BATTERY CHARGER
The charger is not intended to be serviced. When unusable it can be thrown
away according to local waste disposal regulations.
RECALIBRATING MC6
As any test and measurement device, we recommend to recalibrate MC6 at
regular intervals. This ensures accuracy and operating reliability throughout
MC6's lifetime.
MC6 is a high-accuracy calibrator and it should be recalibrated only at laborato-
ries being able to offer sufficient uncertainty. Please note that all calibration
laboratories are not able to offer required uncertainty. It is highly recommended
that the calibration laboratory is accredited (ISO 17025) and has certified quality
system in place.
For advice, please contact Beamex.
By sending the calibrator to Beamex or your closest
Beamex authorized service center for recalibration,
you can be sure that your equipment will be cleaned,
fully tested, loaded with the latest firmware/features,
and recalibrated in a state-of-the-art calibration facility
that meets the specifications of the calibrator.
Note.
When setting MC6's calibration date, see Part 8, Settings.
If MC6 needs cleaning, use cloth soaked with a mild solution of tall oil soap
(pine soap). Wait a few minutes and then clean using a cloth moistened with
pure water. Never use any strong detergents.
Note.
To clean the display, use a microfiber cloth. If necessary, use a
mild detergent and remove it thoroughly when ready.