Safety - Safety Precautions and Warnings 91
SAFETY PRECAUTIONS AND WARNINGS
MC6 is a precision tool that should be used by skilled people who have read
and understood this manual. Working with MC6 involves the usage of pressure
and/or electrical instruments. Be sure to know how to work with these instru-
ments and how to safely connect/disconnect pressure hoses as well as electri-
cal test leads clips, etc.
Sometimes it is necessary to use a portable radio transceiver while working with
the calibrator. To prevent calibration errors caused by the radio frequency inter-
ference, keep the radio far (at least 1 meter) from the calibrator and the circuit
under calibration while sending.
Use MC6 only if you are certain of that it can be used
safely. Safe use of MC6 is no longer possible if one or
more of the following cases are true:
When the case of MC6 is evidently damaged
When MC6 is not functioning as expected
After prolonged storage in unfavorable condi-
tions
After serious damage during transport
Temperature -10 ... 45 °C (14 ... 113 °F)
Humidity 0 to 80 % R.H. non condensing
Max. altitude 3000 m (approx. 9,800 ft)
See also storing conditions in chapter: Storing, on page 93 and charging condi-
tions in chapter Charging the Lithium Polymer Battery Pack, on page 93.
Do not use MC6 in any other way than as described in this User Manual. If this equip-
ment is used in a manner not specified by the manufacturer, the protection provided by
it against hazards may be impaired.
Do not tap the touch screen with sharp or hard objects or
press hard on the display, especially with fingernails. In-
stead, tap lightly using your fingertips.