EasyManua.ls Logo

BEAMEX MC6 - Doing a Data Log - Configuring

BEAMEX MC6
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Group Calibration - Calibrating a Group 47
GROUP SETTINGS
In Group Settings you can define two things:
Calibration Point Sequence, i.e. how the calibration points are run
through. More details below.
When the Instrument Overview (Instrument Info) window appears dur-
ing Group Calibration.
Calibration Point Sequence options:
By Group: Suitable when all instruments selected into the group have
the same input span. The instruments may have different amount of cal-
ibration points as long as the input span is the same.
By Value: The instruments selected into the group may have a span of
any size. The only important thing is: The group's first instrument has
to have the smallest/lowest calibration point of all instruments in
the group. When necessary, use the sorting tool presented in chapter
Editing a Group on page 46.
PERFORMING THE CALIBRATION
The actual calibration is done from point to point similarly as a non-Group Cali-
bration. You may change the default calibration order of Instruments/Functions
by tapping the Info button (see adjacent picture) and selecting another Instru-
ment/Function by browsing the Instrument Overview window.
After all points of all Instruments/Functions are done, the calibration result pag-
es open one by one and you have the chance to save (or not) the calibration
results for each Instrument/Function in the Group.
When the results of all Instruments/Functions are saved (or not), MC6 returns to
the Calibration window. Now you may do another calibration run or return to
the instrument list to, e.g. remove Instruments/Functions from the Group that
did pass in the As Found test and leave the Instruments/Functions in the Group
that require As Left calibration.
Group Settings Window.
Note.
Remember to change connections (if necessary) when the In-
strument/Function changes during the Group calibration.
Info button highlighted.
(2
nd
instrument of a group of three @ 75 °C)
Hint.
Examples of Instruments/Functions suited for Group Calibra-
tion:
• A measurement loop, including a temperature transmitter,
a local temperature indicator and a temperature indicator
in the control room.
• A set of temperature sensors calibrated simultaneously
using, e.g. a dry block.

Table of Contents

Other manuals for BEAMEX MC6

Related product manuals