6-2 THYROID UPTAKE Biodex Medical Systems, Inc. © 2014
NOTE: If you have not yet calibrated your Atomlab 960 on this day, the system will
prompt you to calibrate before proceeding with the Thyroid Uptake or any other test
procedure selected. High Voltage must also be set before using the system for the first
time. Once the calibration and High Voltage have been performed, you may proceed.
WARNING: You must use the same geometry for counting the sample/capsule or lab
background as used for counting the thyroid and / patient background. The system
calculation assumes the same geometry when comparing counts to counts. You must
also ensure that the collimator is at a right angle to the neck phantom and the thyroid
so that it is positioned correctly for counting to retain the correct geometry. Make sure
the distance bar and positioning light are not in the counting field.
AVERTISSEMENT: Vous devez utiliser la même géométrie de comptage de
l'échantillon/capsule ou lab arrière-plan tel qu'utilisé pour le comptage de la thyroïde et
/ patient arrière-plan. Le calcul de système suppose la géométrie même lorsque l'on
compare le nombre de chefs d'accusation. Vous devez également vous assurer que le
collimateur est à angle droit avec le fantôme de cou et de la thyroïde afin qu'il soit
positionné correctement pour le comptage pour conserver la géométrie correcte.
Assurez-vous que la barre d'écartement et positionnement de la lumière ne sont pas
dans le champ de comptage.
Performing A Thyroid Uptake
(See Figure 6.2.)
Figure 6.2. The Select or Add a Patient screen.