C. APPENDIX C –
ELECTROMAGNETIC COMPATIBILITY
This equipment conforms to the following Safety Standards:
Table 1.1 Safety standards
ACCOMPANYING EMC DOCUMENTS
This medical electrical equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Use of accessories, transducers and cables other than those specified, with the exception of
accessories, transducers and cables sold by the manufacturer of this equipment, as
replacement parts for internal and external components, may result in increased emissions or
decreased immunity of the equipment.
The Atomlab 960 Thyroid Uptake System should not be used adjacent to or stacked with
other equipment. If the Atomlab 960 Thyroid Uptake System is used while positioned adjacent
to other equipment, it should be observed to verify normal operation in the configuration in
which it will be used.
LIST OF CABLE ACCESSORIES
The list in Table 1.2 includes all accessory cables supplied with the Atomlab 960 Thyroid
Uptake System for which the manufacturer of this equipment claims compliance to
EN 60601-1-2 when used with the Atomlab 960 Thyroid Uptake System.
Table 1.2 Atomlab 960 Thyroid Uptake System cables