B-2 APPENDIX B Biodex Medical Systems, Inc. © 2014
Optional:
187-602 Well Counter:
Detector: 2" x 2" Nal (Tl) integral line scintillation detector with a.75" diameter x 1.44" deep well
(1.9 x 3.7 cm)
Lead Shielding: 1" thick (2.5 cm)
Cover: .125" thick (.32 cm)
Weight: 50 lb (22.7 kg)
187-603 Lead Shield, Well Counter
Lead Shielding: 1" thick (5 cm)
Weight: 69 lb (31.4 kg)
Power: Low voltage transformer with hospital grade power cord and plug.
Line Voltage:
120 VAC, 60 Hz, 2 AMP circuit breaker
230 VAC, 50 Hz, 1 AMP circuit breaker
Certification: ETL Listed to UL 60601-1 Std. and CAN CSA C22.2 No. 601.1-M90, 2nd and 3rd
editions, and CE marked.
Atomlab Line Cord:
The line cord for the AL 960 is non-detachable. If it should become damaged, a qualified
service technician should be called to repair or replace it.
The following describes the procedure for replacing the line cord on the AL960.
1. Unplug the AL960 line cord from the wall receptacle.
2. Remove the transformer cover plate.
3. Remove the earth ground from the chassis ground stud.
4. Remove the line and neutral conductors from the line side of the circuit breaker.
5. Loosen the gland nut on the heyco strain relief located where the line cord enters the
AL960 cart.
6. Carefully pull the old line cord out of the transformer compartment and through the heyco
in the AL 960 chassis.
7. Reverse the order of these instructions and reassemble with new power cord.
Certification: ETL Listed to UL 60601-1 Std. and CAN CSA C22.2 No. 601.1-M90, 2
nd
and 3
rd
editions, and CE marked.