5-4 QUALITY ASSURANCE Biodex Medical Systems, Inc. © 2014
High Voltage Adjustment
(See Figure 5.5.)
Figure 5.5. The MCA Operation In Progress screen is displayed while high voltage measurement is being performed.
High Voltage Adjustment should be performed the very first time the system is used, if the
system is to be used in a new environment, or in the instance where a substantial change has
been made, i.e., when using a new tube.
NOTE: If a probe and well are both used the high voltage adjustment must be set independently
for each detector. The system must be ON and warmed-up prior to setting the high voltage.